Fruit fly (adult) Bullet Blender® homogenization protocol
Extract molecules (DNA, RNA, protein, chemicals)
Wet final product
Sample sizes: 100 to 1000 mg.
Notes on the protocol: This protocol does not specify a particular buffer - you may choose which is most appropriate for your downstream application (nucleic acid isolation, protein extraction, etc.).
Materials Required
uL
One of these Bullet Blenders
Bullet Blender 5 Storm (BBY5M)
Bullet Blender 5E (BBY5E)
Bullet Blender 5 Gold (BB5E-AU)
Reagents
Homogenization buffer
2 x volume of sample
PBS (optional)
2 x volume of sample
Bead choices
0.5 mm zirconium oxide beads (ZROB05) Use a volume of beads equivalent to 1 x the volume of the sample
Procedure
Place the sample in the tube with the beads.
(Optional) Wash the sample 3x with 1/2 tube volume of PBS to remove surface contaminants.
Close the tubes tightly and place them in the Bullet Blender.
Set the controls for Speed 8 and Time 3. Press Start.
After the run, remove the tubes from the instrument and visually inspect the samples. If homogenization is incomplete, repeat the homogenization step at a higher speed.