Fruit tree twigs Bullet Blender® homogenization protocol
Extract molecules (DNA, RNA, protein, chemicals)
Wet final product
Sample sizes: 50 to 150 mg.
Notes on the protocol: This protocol does not specify a particular buffer - you may choose which is most appropriate for your downstream application (nucleic acid isolation, protein extraction, etc.).
Materials Required
uL
One of these Bullet Blenders
Bullet Blender 5E (BBY5E)
Bullet Blender 5 Gold (BB5E-AU)
Reagents
Homogenization buffer
500 ul
Bead choices
5.6 mm stainless steel UFO beads (SSUFO56) use 6 beads
Procedure
Select a 3-4 cm. long section of stem. Cross-cut into pieces approx. 1 cm. long
Add the sample, buffer and beads to the tube in the amounts shown above. Sample volume may be approximated by sample weight.
Set the controls for Speed 12 and Time 5. Press Start.
After the run, remove the tubes from the instrument and visually inspect the samples. If homogenization is incomplete, repeat the homogenization step at a higher speed.