Notes on the protocol: This protocol does not specify a particular buffer - you may choose which is most appropriate for your downstream application (nucleic acid isolation, protein extraction, etc.).
Materials Required
uL
One of these Bullet Blenders
Bullet Blender 50-DX (BB50-DX)
Bullet Blender 50 Gold (BB50-AU)
Reagents
Homogenization buffer
2 x volume of sample
Bead choices
0.9 - 2.0 mm stainless steel blend (SSB14B) Use a volume of beads equivalent to 1 x the volume of the sample
Procedure
Place the sample in the tube with the beads.
Add a volume of buffer that is twice the volume of the sample. Sample volume may be approximated by sample weight. E.g., for a 100 mg. sample, add 0.2 ml. buffer.
Close the tubes tightly and place them in the Bullet Blender.
Set the controls for Speed 8 and Time 12. Press Start.
After the run, remove the tubes from the instrument and visually inspect the samples. If homogenization is incomplete, repeat the homogenization step at a higher speed.